4.3 Inductively coupled plasma mass spectrometry (ICP-MS)
10 min read•august 21, 2024
() is a powerful tool for analyzing elements and isotopes in geological samples. It combines high-temperature plasma to ionize samples with mass spectrometry to measure ions, allowing precise detection of trace elements and isotope ratios.
ICP-MS has revolutionized geochemical analysis with its , multi-element capabilities, and ability to measure isotope ratios. Understanding its principles, instrumentation, and data processing is key for geochemists to unlock insights into Earth's composition and processes.
Principles of ICP-MS
Inductively Coupled Plasma Mass Spectrometry (ICP-MS) revolutionized elemental analysis in geochemistry by enabling precise measurement of trace elements and isotope ratios
ICP-MS combines high-temperature plasma ionization with mass spectrometry to analyze complex geological samples, providing insights into Earth's composition and processes
Plasma generation process
Top images from around the web for Plasma generation process
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
1 of 2
Top images from around the web for Plasma generation process
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
GChron - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) U–Pb carbonate ... View original
Is this image relevant?
1 of 2
Utilizes argon gas flowing through concentric quartz tubes
Radio frequency (RF) power applied to a copper coil creates oscillating electromagnetic field
Seed electrons from a spark initiator collide with argon atoms, generating ions and more electrons
Cascade effect produces sustained plasma reaching temperatures up to 10,000 K
High temperature efficiently atomizes and ionizes sample material
Ion formation mechanisms
Sample aerosol introduced into plasma undergoes desolvation, vaporization, atomization, and ionization
Thermal ionization occurs as atoms absorb energy from plasma, ejecting electrons
Penning ionization involves energy transfer from metastable argon atoms to analyte atoms
Charge transfer reactions between argon ions and sample atoms produce analyte ions
Majority of elements ionized to singly-charged positive ions (M+)
Mass spectrometry basics
Separates ions based on their mass-to-charge ratio (m/z)
Ions accelerated and focused into a beam using electrostatic lenses
Mass analyzer (quadrupole, magnetic sector, or time-of-flight) separates ions
Detector measures ion signal intensity, typically using an electron multiplier
Mass spectrum produced shows ion intensity vs m/z, allowing quantitative analysis
ICP-MS instrumentation
ICP-MS instruments consist of several integrated components working together to analyze samples
Understanding each component's function crucial for optimizing performance and interpreting results in geochemical applications
Sample introduction systems
Nebulizer converts liquid sample into fine aerosol
Pneumatic nebulizers use high-velocity gas flow (concentric, cross-flow designs)
Ultrasonic nebulizers improve efficiency for dilute samples
Spray chamber removes large droplets, ensuring uniform aerosol reaches plasma
Desolvation systems reduce solvent load, improving sensitivity for some applications
Plasma torch design
Consists of three concentric quartz tubes: outer, intermediate, and sample injector
Outer tube carries coolant gas (typically 12-17 L/min) to protect torch from melting
Intermediate tube carries auxiliary gas (0.5-1.5 L/min) to keep plasma away from injector tip
Sample injector introduces aerosol into plasma core (0.8-1.2 L/min carrier gas flow)
RF coil surrounds torch, typically operating at 27 or 40 MHz
Ion optics and focusing
Extract ions from atmospheric pressure plasma into vacuum system
Sampler cone (~ 1 mm orifice) allows ions to enter first vacuum stage
Skimmer cone (~ 0.4-0.8 mm orifice) further focuses ion beam
Electrostatic lenses focus and steer ion beam towards mass analyzer
Photon stop or shadow stop blocks neutral species and photons
Mass analyzers
Quadrupole mass filters most common due to speed and simplicity
Four parallel rods with applied DC and RF voltages create oscillating electric field
Only ions with specific m/z have stable trajectories and pass through quadrupole
Magnetic sector analyzers offer higher resolution but are more expensive
Time-of-flight analyzers provide rapid, simultaneous multi-element detection
Detector types
Electron multiplier most common, converts ion impacts into measurable electrical signal
Discrete dynode electron multipliers use series of dynodes to amplify signal
Continuous dynode (channeltron) multipliers have curved tube design
Faraday cup detectors used for high ion currents or precise isotope ratio measurements
Dual mode detectors combine electron multiplier and Faraday cup for wide dynamic range
Sample preparation techniques
Proper sample preparation critical for accurate and precise ICP-MS analysis in geochemistry
Techniques aim to create homogeneous solutions while minimizing contamination and matrix effects
Dissolution methods
Acid digestion common for silicate rocks and minerals (HF, HNO3, HCl mixtures)
Microwave-assisted digestion speeds up process and reduces contamination risk
Fusion techniques (lithium metaborate, sodium peroxide) for refractory minerals
Aqua regia digestion for partial extraction of metals from sulfides and some oxides
Specialized methods for organic-rich samples (H2O2 addition, high-pressure ashing)
Dilution strategies
Serial dilutions used to bring sample concentrations within calibration range
Matrix matching involves diluting samples and standards to similar total dissolved solids
Gravimetric dilution provides higher precision than volumetric methods
Internal standards added to compensate for matrix effects and instrument drift
Online dilution systems allow automated, real-time sample dilution
Matrix effects mitigation
Standard addition method accounts for matrix-induced signal suppression or enhancement
Matrix separation techniques (ion exchange, solvent extraction) remove interfering elements
High dilution factors reduce matrix effects but may compromise detection limits
Collision/reaction cells in ICP-MS can reduce polyatomic interferences
Mathematical corrections applied based on known interference patterns
Analytical capabilities
ICP-MS offers exceptional analytical performance for geochemical applications
Understanding instrument capabilities essential for method development and data interpretation
Detection limits vs sensitivity
Detection limits typically in parts per trillion (ppt) to parts per quadrillion (ppq) range
Sensitivity defined as signal intensity per unit concentration (counts per second per ppb)
Factors affecting detection limits include background noise, matrix effects, and interferences
High-resolution ICP-MS improves detection limits for interfered elements
Sensitivity varies across mass range due to mass bias effects
Precision and accuracy
Precision typically 1-3% RSD for most elements at moderate concentrations
Accuracy depends on calibration quality, interference corrections, and matrix matching
Isotope ratio measurements can achieve precision better than 0.1% RSD
Long-term stability affected by factors like cone condition and plasma stability
Reference materials crucial for assessing and demonstrating accuracy
Multi-element analysis
Simultaneous or rapid sequential analysis of 20-30 elements common
Full mass scans (2-260 amu) possible in seconds to minutes
Semiquantitative analysis mode for rapid screening of unknown samples
Dynamic range spans up to 9 orders of magnitude using pulse-counting and analog detection
Ability to measure major, minor, and trace elements in a single analysis
Isotope ratio measurements
Precise measurement of isotope ratios for geochemical tracers and
Applications include Sr, Nd, Pb isotope systems for petrogenesis and provenance studies
U-Pb dating of zircons and other minerals for age determination
Stable isotope ratio analysis (e.g., Li, B, Fe) for process tracing
Mass bias corrections applied using internal normalization or external bracketing
Interferences in ICP-MS
Interferences pose significant challenges in ICP-MS analysis of geological samples
Understanding and mitigating interferences crucial for accurate elemental and isotopic measurements
Spectral interferences
Isobaric interferences occur when different elements have isotopes of the same nominal mass
Polyatomic interferences form from combinations of plasma gas, matrix, and solvent species
Common polyatomics include ArO+, ArAr+, and oxide species (MO+)
Doubly-charged ions (M2+) appear at half their true mass, interfering with other elements
High-resolution instruments can resolve some spectral interferences
Non-spectral interferences
Matrix effects cause signal suppression or enhancement due to sample composition
Space-charge effects in the ion beam affect light elements more than heavy elements
Memory effects result from carryover between samples, especially for certain elements (B, Hg)
Physical interferences from high dissolved solids can clog nebulizer or deposit on cones
Ionization suppression in the plasma due to easily ionized elements (Na, K)
Interference correction methods
Mathematical corrections based on natural isotope abundances and interference formation rates
Cool plasma conditions reduce formation of some argon-based interferences
Collision/reaction cells use collision gases (He) or reactive gases (H2, NH3, O2) to remove interferences
Chemical separations to remove interfering elements prior to analysis
Isotope pattern deconvolution for resolving complex interference scenarios
Applications in geochemistry
ICP-MS versatility makes it indispensable for various geochemical investigations
Ability to analyze diverse sample types provides insights into Earth processes across multiple scales
Trace element analysis
Determination of rare earth elements (REE) patterns for petrogenetic studies
Transition metal concentrations in minerals for understanding ore formation processes
Chalcophile element distributions in magmatic systems to trace sulfide saturation
Fluid-mobile element concentrations in metamorphic rocks to study metasomatism
Trace metal analysis in environmental samples for pollution monitoring
Isotope fingerprinting
Sr-Nd-Pb isotope systematics to determine magma sources and crustal contamination
Hf isotopes in zircons to trace crustal evolution and recycling
Os isotopes in mantle-derived rocks to study core-mantle interactions
Cu and Zn isotopes in ore deposits to understand metal transport and precipitation
B and Li isotopes as tracers of fluid-rock interactions and weathering processes
Geochronology applications
U-Pb dating of zircons, monazites, and other accessory minerals
Re-Os dating of sulfides and organic-rich sediments
Lu-Hf dating of garnet for metamorphic chronology
U-series disequilibrium dating of young volcanic rocks
Trace element thermochronology (e.g., Zr-in-rutile) for thermal history reconstruction
Environmental monitoring
Heavy metal contamination assessment in soils, sediments, and waters
Rare earth element patterns as tracers of sediment provenance
Isotope ratio analysis to fingerprint sources of atmospheric particulates
Biomonitoring using trace element concentrations in plants and animals
Water quality analysis for both dissolved and particulate trace elements
Data processing and interpretation
Raw ICP-MS data requires careful processing and interpretation to extract meaningful geochemical information
Understanding data reduction techniques essential for producing high-quality results
Calibration methods
External calibration using multi-element standard solutions
Standard addition method for complex matrices with significant interferences
Isotope dilution for highest accuracy in concentration and isotope ratio measurements
Semi-quantitative calibration using full mass scans and theoretical response factors
Matrix-matched calibration to account for matrix effects in specific sample types
Internal standardization
Addition of elements not present in sample to correct for matrix effects and instrument drift
Common internal standards include In, Rh, and Bi for different mass ranges
Multiple internal standards can be used to correct mass-dependent effects
Online addition of internal standards ensures consistent spike concentrations
Selection of appropriate internal standards based on ionization potential and mass
Data reduction techniques
Background subtraction to remove contributions from reagent blanks and instrument noise
Interference corrections applied based on measured intensities of monitor isotopes
Mass bias correction for accurate isotope ratio measurements
Drift correction using periodic measurements of quality control standards
Propagation of uncertainties from counting statistics, calibration, and corrections
Quality control measures
Regular analysis of certified reference materials to assess accuracy and precision
Method blanks to quantify contamination from reagents and sample preparation
Duplicate analyses to evaluate reproducibility
Spike recovery tests to check for matrix effects and interferences
Long-term monitoring of instrument sensitivity and stability using control charts
Advantages and limitations
Understanding strengths and weaknesses of ICP-MS crucial for selecting appropriate analytical techniques
Comparison with other methods helps optimize geochemical research strategies
ICP-MS vs other techniques
Superior detection limits compared to ICP-OES for most elements
Faster multi-element analysis than traditional atomic absorption spectroscopy (AAS)
Better precision for isotope ratios than thermal ionization mass spectrometry (TIMS) for some systems
More versatile than X-ray fluorescence (XRF) for in diverse sample types
Complementary to electron microprobe analysis (EPMA) for bulk vs. in-situ measurements
Sample consumption considerations
Typically requires 1-5 mL of solution per analysis, more efficient than some other techniques
ICP-MS allows for micro-sampling with minimal sample destruction
Improves detection limits for interfered elements (Fe, K, Ca, As, Se)
Enables accurate analysis of traditionally difficult elements in geochemical samples
Key Terms to Review (28)
Calibration Curve: A calibration curve is a graphical representation that illustrates the relationship between known concentrations of a substance and the response produced by an analytical method. This curve is essential for quantifying unknown samples by comparing their response to the established relationship, ensuring accuracy and reliability in measurements, which is critical in various applications, such as isotope standards, mass spectrometry, radiometric dating, and ICP-MS techniques.
Delta Notation: Delta notation is a way of expressing the relative difference in the isotopic composition of a sample compared to a standard reference material. This notation helps in understanding variations in isotopic abundances, crucial for analyzing atomic structure and isotopes, evaluating isotope effects in equilibria, and interpreting results from mass spectrometry techniques.
Detection limit: The detection limit is the smallest concentration of an analyte that can be reliably detected but not necessarily quantified under the stated experimental conditions. This concept is crucial in analytical chemistry, as it defines the threshold below which measurements are considered unreliable and helps in assessing the sensitivity of techniques like inductively coupled plasma mass spectrometry.
Dilution strategies: Dilution strategies refer to methods used to reduce the concentration of a sample solution before analysis, particularly in mass spectrometry. This process is crucial in inductively coupled plasma mass spectrometry (ICP-MS) as it helps prevent signal saturation and matrix effects that can interfere with accurate measurements of trace elements and isotopes.
Dissolution methods: Dissolution methods are techniques used to dissolve solid samples into liquid forms, enabling their analysis through various analytical techniques. These methods are critical in isotope geochemistry, as they facilitate the release of analytes from solid matrices, allowing for accurate measurements and characterizations. The choice of dissolution method can significantly affect the quality and reliability of the data obtained from subsequent analyses.
Environmental Monitoring: Environmental monitoring refers to the systematic collection, analysis, and interpretation of data related to environmental conditions, aimed at assessing the health of ecosystems and detecting changes over time. This process helps identify potential pollution sources, track changes in natural resources, and evaluate the effectiveness of environmental policies and regulations. The integration of advanced analytical techniques enhances the ability to monitor environmental quality and safety.
Frederick A. A. Miller: Frederick A. A. Miller is a prominent figure in the field of geochemistry, specifically known for his contributions to the development of inductively coupled plasma mass spectrometry (ICP-MS). His work has significantly advanced the analytical techniques used to measure isotopic compositions and trace elements in geological samples, enhancing our understanding of Earth's processes and material properties.
Geochronology: Geochronology is the science of determining the age of rocks, fossils, and sediments through the study of their isotopes and radioactive decay processes. This field plays a critical role in understanding the timing of geological events, the history of the Earth, and the processes involved in crustal growth and recycling.
Geochronology applications: Geochronology applications refer to the use of various dating techniques to determine the age of rocks, minerals, and sediments. This field is crucial for understanding geological history, including the timing of events such as volcanic eruptions, sediment deposition, and tectonic movements. It often employs methods such as radiometric dating to provide insights into the Earth's formation and evolution.
ICP-MS: Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is an analytical technique used to detect and quantify trace elements in a sample by measuring the ions generated in an inductively coupled plasma. It combines the high sensitivity of mass spectrometry with the ability to handle various sample types, making it essential for geochemical analyses and environmental monitoring.
Inductively Coupled Plasma Mass Spectrometry: Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is a powerful analytical technique used to detect and quantify trace elements and isotopes in various samples. It works by ionizing the sample with an inductively coupled plasma and then analyzing the ions with mass spectrometry, making it essential for determining isotopic ratios, understanding radiometric dating, and assessing environmental contamination.
Interference correction methods: Interference correction methods are techniques used in analytical chemistry to account for and correct measurement inaccuracies caused by overlapping signals from different isotopes or ions in mass spectrometry. These methods ensure that the results obtained from instruments like inductively coupled plasma mass spectrometry (ICP-MS) accurately reflect the concentration of the target analytes, rather than being skewed by the presence of interferences that can arise during the measurement process.
Isotope fingerprinting: Isotope fingerprinting is a technique used to identify the unique isotopic signatures of elements in a sample, which can reveal information about its origin, history, and processes it has undergone. This method leverages the variations in the isotopic composition of elements to trace sources of materials, assess environmental changes, or study biogeochemical cycles.
Isotope fractionation: Isotope fractionation is the process that leads to the separation of isotopes of an element due to physical or chemical processes, resulting in a variation of their ratios in different substances. This phenomenon is critical for understanding various natural processes, as it influences the isotopic composition of elements in geological, environmental, and extraterrestrial contexts. The concept helps in interpreting delta values, analyzing materials with advanced mass spectrometry techniques, and assessing the impact of contamination in groundwater or the composition of lunar samples.
Isotopic Signature: An isotopic signature is a unique set of isotopic ratios that can provide insights into the source, processes, or history of a material. This signature can reveal information about the environment, biological processes, or chemical interactions, making it a powerful tool in various scientific fields. By analyzing isotopic signatures, scientists can trace origins, assess changes over time, and differentiate between similar materials based on their distinct isotopic compositions.
Laser ablation: Laser ablation is a material removal process that uses focused laser energy to vaporize or remove material from a solid surface. This technique is crucial in geochemical analysis, particularly for precise sampling and analysis of solid materials, allowing for the detailed study of isotope compositions in various geological contexts.
Mass discrimination: Mass discrimination refers to the differential mass-dependent behavior of isotopes during physical processes, which can lead to variations in isotopic composition. This phenomenon is crucial in fields like isotope geochemistry and analytical techniques, where understanding the separation of isotopes is key for interpreting data from various samples.
Mass spectrometer: A mass spectrometer is an analytical instrument used to measure the mass-to-charge ratio of ions, allowing for the identification and quantification of chemical species based on their mass. This technology plays a crucial role in isotope geochemistry by enabling the detection and analysis of isotopes in various materials, facilitating the understanding of atomic structure, isotopic compositions, and elemental concentrations in samples through different techniques.
Matrix effects mitigation: Matrix effects mitigation refers to strategies employed to reduce or eliminate the interference caused by co-existing substances in a sample that can affect the accuracy of analytical measurements. This is particularly critical in techniques like inductively coupled plasma mass spectrometry (ICP-MS), where complex sample matrices can lead to signal suppression or enhancement, resulting in inaccurate quantification of target analytes.
Paleoceanography: Paleoceanography is the study of the history and changes in the ocean's properties, chemistry, and biological activity over geological time. This field helps scientists understand past climate conditions, ocean circulation patterns, and how marine life adapted to changes in the environment. By examining various indicators, paleoceanography reveals insights into Earth's climate history, including how stable isotope ratios can indicate past temperatures and salinity levels, while also connecting with biogeochemical cycles such as the oxygen cycle.
Plasma source: A plasma source is a device or system that generates a high-temperature, ionized gas known as plasma, which is essential for various analytical techniques, including mass spectrometry. In the context of inductively coupled plasma mass spectrometry (ICP-MS), the plasma source facilitates the ionization of sample atoms, allowing for their subsequent detection and quantification. This process is crucial for achieving high sensitivity and precision in elemental analysis.
Radioactive Isotopes: Radioactive isotopes are variants of chemical elements that have an unstable nucleus and emit radiation as they decay into more stable forms. This process can involve the release of particles or electromagnetic radiation, leading to a change in the element's identity over time. These isotopes are crucial for understanding nuclear stability, processes in nature, and analytical techniques used in geochemistry.
Robert J. Walker: Robert J. Walker was a significant figure in the development of inductively coupled plasma mass spectrometry (ICP-MS), known for his contributions to advancing analytical techniques used in geochemistry and other scientific fields. His work laid the groundwork for utilizing ICP-MS as a powerful tool for precise isotopic and elemental analysis, enhancing the capability to measure trace elements in various samples.
Sample digestion: Sample digestion is a process used to break down complex materials into simpler, more analyzable components, often through chemical reactions. This technique is essential for preparing samples for analytical methods, ensuring that the elements of interest are in a suitable form for accurate measurement and quantification.
Sample preparation techniques: Sample preparation techniques are methods used to prepare materials or samples for analysis in various scientific fields. These techniques ensure that samples are in a suitable form for accurate measurements and help to eliminate potential interferences that could affect the results, particularly in sophisticated analytical methods like inductively coupled plasma mass spectrometry (ICP-MS). Proper sample preparation is crucial for achieving reliable and reproducible data.
Sensitivity: Sensitivity in the context of inductively coupled plasma mass spectrometry (ICP-MS) refers to the instrument's ability to detect and quantify low concentrations of elements in a sample. It plays a crucial role in determining the minimum detectable limits of various isotopes, impacting the precision and accuracy of analytical results. High sensitivity ensures that trace elements can be identified, which is essential for applications in environmental monitoring, geochemistry, and health sciences.
Stable Isotopes: Stable isotopes are variants of chemical elements that have the same number of protons but a different number of neutrons, resulting in no radioactive decay over time. These isotopes are important for understanding various geological, environmental, and biological processes, as their abundances can provide insights into everything from ancient climate conditions to the origins of planetary bodies.
Trace Element Analysis: Trace element analysis is the process of detecting and quantifying the presence of trace elements in various materials, often at extremely low concentrations. This type of analysis is crucial for understanding geochemical processes, environmental monitoring, and assessing the quality of natural resources. It helps scientists and researchers uncover information about the composition and history of geological samples, minerals, and even biological systems.