🩺Biomedical Instrumentation Unit 5 – Bioelectric Signals and ECG
Bioelectric signals are the foundation of many bodily functions, originating from cellular electrical activity. These signals, including action potentials, are crucial for understanding physiological processes and can be measured using various electrode types.
The electrocardiogram (ECG) is a prime example of bioelectric signal application, recording the heart's electrical activity. It provides vital information about cardiac function, including heart rate, rhythm, and potential abnormalities, making it an essential tool in clinical diagnostics and patient monitoring.
Bioelectric signals originate from the electrical activity of cells and tissues in the body
Generated by the movement of ions across cell membranes, creating potential differences
Cells involved in bioelectric signal generation include neurons, muscle cells, and endocrine cells
Action potentials are the primary mechanism for signal transmission in excitable cells
Occur when the cell membrane depolarizes and repolarizes rapidly
Triggered by stimuli that exceed the cell's threshold potential
Bioelectric signals can be measured using electrodes placed on the skin surface (surface electrodes) or within the body (intracellular or extracellular electrodes)
Signal characteristics depend on the type of tissue, location, and physiological state
Bioelectric signals are typically low in amplitude (microvolts to millivolts) and require amplification for analysis
Signal processing techniques are used to filter, analyze, and interpret bioelectric signals (Fourier analysis, wavelet analysis)
Physiological Basis of ECG
ECG (electrocardiogram) records the electrical activity of the heart over time
Cardiac cells generate action potentials that propagate through the heart, causing contraction
The conduction system of the heart coordinates the sequence of electrical activation
Sinoatrial (SA) node initiates the heartbeat, acting as the natural pacemaker
Atrioventricular (AV) node delays the impulse, allowing time for ventricular filling
His-Purkinje system rapidly distributes the impulse to the ventricles
Each phase of the cardiac cycle produces a characteristic waveform on the ECG
P wave represents atrial depolarization
QRS complex represents ventricular depolarization
T wave represents ventricular repolarization
Changes in ECG waveforms can indicate various cardiac abnormalities (arrhythmias, ischemia, infarction)
ECG provides valuable information about heart rate, rhythm, and conduction
ECG Signal Characteristics
ECG signal consists of a series of waves and intervals that represent different phases of the cardiac cycle
Key components of the ECG signal include P wave, QRS complex, T wave, and U wave (not always visible)
Amplitude of ECG waveforms is typically measured in millivolts (mV)
P wave amplitude: 0.1-0.2 mV
QRS complex amplitude: 1-2 mV
T wave amplitude: 0.2-0.5 mV
Duration of ECG waveforms and intervals is measured in milliseconds (ms)
P wave duration: 60-120 ms
QRS complex duration: 60-100 ms
T wave duration: 100-250 ms
ECG signal frequency content ranges from 0.05 to 100 Hz, with most diagnostic information in the 0.5-50 Hz range
Signal-to-noise ratio (SNR) is an important consideration in ECG signal quality
Artifacts and noise can distort the ECG signal (muscle activity, electrode movement, power line interference)
ECG Measurement Techniques
Standard 12-lead ECG is the most common measurement technique
Uses 10 electrodes placed on the limbs and chest to record 12 different views of the heart's electrical activity
Leads I, II, and III are bipolar limb leads that measure potential differences between electrodes
Leads aVR, aVL, and aVF are augmented unipolar limb leads that measure potential differences relative to a reference point
Leads V1-V6 are unipolar chest leads that measure potential differences relative to a reference point
Holter monitoring is used for continuous ECG recording over an extended period (24-48 hours)
Portable device worn by the patient to record ECG during daily activities
Useful for detecting intermittent arrhythmias or ischemic events
Stress ECG (exercise stress test) measures the heart's response to physical exertion
Patient undergoes graded exercise on a treadmill or stationary bike while ECG is recorded
Helps diagnose coronary artery disease and assess exercise capacity
Intracardiac electrogram (EGM) records electrical activity directly from the heart using catheters or implantable devices
Provides more detailed and localized information than surface ECG
Used during electrophysiology studies and ablation procedures
ECG Instrumentation and Equipment
ECG instrumentation consists of electrodes, lead wires, amplifiers, filters, and a display or recording device
Electrodes transduce ionic currents in the body into electrical currents in the measurement circuit
Types of electrodes include disposable adhesive electrodes, reusable disc electrodes, and needle electrodes
Electrode-skin interface is a critical factor in signal quality and requires proper skin preparation and electrode placement
Lead wires connect the electrodes to the ECG amplifier and should be shielded to minimize interference
ECG amplifiers boost the low-amplitude bioelectric signals to a level suitable for further processing and display
Differential amplifiers are used to reject common-mode noise and amplify the difference between two input signals
Typical ECG amplifier gain is 1000-5000, with a bandwidth of 0.05-100 Hz
Filters remove unwanted frequency components from the ECG signal
Low-pass filters remove high-frequency noise and artifacts (50-100 Hz cutoff)
Notch filters remove power line interference (50 or 60 Hz)
Analog-to-digital converters (ADCs) sample and digitize the analog ECG signal for digital storage, processing, and analysis
Sampling rates of 250-1000 Hz are commonly used in ECG acquisition
Resolution of 12-16 bits is typical for ECG ADCs
Signal Processing in ECG
Signal processing techniques are used to improve the quality and interpretability of ECG signals
Filtering is used to remove noise and artifacts from the ECG signal
Digital filters (FIR, IIR) are implemented in software for more flexibility and control
Adaptive filters can adjust their parameters based on signal characteristics
Baseline wander correction removes low-frequency drift in the ECG signal
High-pass filtering, polynomial fitting, or cubic spline interpolation can be used
Powerline interference removal eliminates 50 or 60 Hz noise from the ECG signal
Notch filtering, adaptive filtering, or subtraction of a noise template can be used
QRS detection identifies the location and timing of QRS complexes in the ECG signal
Algorithms based on amplitude, slope, or wavelet transforms are commonly used
Enables the calculation of heart rate and the analysis of rhythm and morphology
Signal averaging improves the signal-to-noise ratio by combining multiple ECG beats
Reduces random noise and enhances small, consistent signal components (P wave, late potentials)
Time-frequency analysis (short-time Fourier transform, wavelet transform) provides information about the time-varying frequency content of the ECG signal
Useful for detecting and characterizing transient events and abnormalities