Biomedical Instrumentation

🩺Biomedical Instrumentation Unit 5 – Bioelectric Signals and ECG

Bioelectric signals are the foundation of many bodily functions, originating from cellular electrical activity. These signals, including action potentials, are crucial for understanding physiological processes and can be measured using various electrode types. The electrocardiogram (ECG) is a prime example of bioelectric signal application, recording the heart's electrical activity. It provides vital information about cardiac function, including heart rate, rhythm, and potential abnormalities, making it an essential tool in clinical diagnostics and patient monitoring.

Fundamentals of Bioelectric Signals

  • Bioelectric signals originate from the electrical activity of cells and tissues in the body
  • Generated by the movement of ions across cell membranes, creating potential differences
  • Cells involved in bioelectric signal generation include neurons, muscle cells, and endocrine cells
  • Action potentials are the primary mechanism for signal transmission in excitable cells
    • Occur when the cell membrane depolarizes and repolarizes rapidly
    • Triggered by stimuli that exceed the cell's threshold potential
  • Bioelectric signals can be measured using electrodes placed on the skin surface (surface electrodes) or within the body (intracellular or extracellular electrodes)
  • Signal characteristics depend on the type of tissue, location, and physiological state
  • Bioelectric signals are typically low in amplitude (microvolts to millivolts) and require amplification for analysis
  • Signal processing techniques are used to filter, analyze, and interpret bioelectric signals (Fourier analysis, wavelet analysis)

Physiological Basis of ECG

  • ECG (electrocardiogram) records the electrical activity of the heart over time
  • Cardiac cells generate action potentials that propagate through the heart, causing contraction
  • The conduction system of the heart coordinates the sequence of electrical activation
    • Sinoatrial (SA) node initiates the heartbeat, acting as the natural pacemaker
    • Atrioventricular (AV) node delays the impulse, allowing time for ventricular filling
    • His-Purkinje system rapidly distributes the impulse to the ventricles
  • Each phase of the cardiac cycle produces a characteristic waveform on the ECG
    • P wave represents atrial depolarization
    • QRS complex represents ventricular depolarization
    • T wave represents ventricular repolarization
  • Changes in ECG waveforms can indicate various cardiac abnormalities (arrhythmias, ischemia, infarction)
  • ECG provides valuable information about heart rate, rhythm, and conduction

ECG Signal Characteristics

  • ECG signal consists of a series of waves and intervals that represent different phases of the cardiac cycle
  • Key components of the ECG signal include P wave, QRS complex, T wave, and U wave (not always visible)
  • Amplitude of ECG waveforms is typically measured in millivolts (mV)
    • P wave amplitude: 0.1-0.2 mV
    • QRS complex amplitude: 1-2 mV
    • T wave amplitude: 0.2-0.5 mV
  • Duration of ECG waveforms and intervals is measured in milliseconds (ms)
    • P wave duration: 60-120 ms
    • QRS complex duration: 60-100 ms
    • T wave duration: 100-250 ms
  • ECG signal frequency content ranges from 0.05 to 100 Hz, with most diagnostic information in the 0.5-50 Hz range
  • Signal-to-noise ratio (SNR) is an important consideration in ECG signal quality
  • Artifacts and noise can distort the ECG signal (muscle activity, electrode movement, power line interference)

ECG Measurement Techniques

  • Standard 12-lead ECG is the most common measurement technique
    • Uses 10 electrodes placed on the limbs and chest to record 12 different views of the heart's electrical activity
    • Leads I, II, and III are bipolar limb leads that measure potential differences between electrodes
    • Leads aVR, aVL, and aVF are augmented unipolar limb leads that measure potential differences relative to a reference point
    • Leads V1-V6 are unipolar chest leads that measure potential differences relative to a reference point
  • Holter monitoring is used for continuous ECG recording over an extended period (24-48 hours)
    • Portable device worn by the patient to record ECG during daily activities
    • Useful for detecting intermittent arrhythmias or ischemic events
  • Stress ECG (exercise stress test) measures the heart's response to physical exertion
    • Patient undergoes graded exercise on a treadmill or stationary bike while ECG is recorded
    • Helps diagnose coronary artery disease and assess exercise capacity
  • Intracardiac electrogram (EGM) records electrical activity directly from the heart using catheters or implantable devices
    • Provides more detailed and localized information than surface ECG
    • Used during electrophysiology studies and ablation procedures

ECG Instrumentation and Equipment

  • ECG instrumentation consists of electrodes, lead wires, amplifiers, filters, and a display or recording device
  • Electrodes transduce ionic currents in the body into electrical currents in the measurement circuit
    • Types of electrodes include disposable adhesive electrodes, reusable disc electrodes, and needle electrodes
    • Electrode-skin interface is a critical factor in signal quality and requires proper skin preparation and electrode placement
  • Lead wires connect the electrodes to the ECG amplifier and should be shielded to minimize interference
  • ECG amplifiers boost the low-amplitude bioelectric signals to a level suitable for further processing and display
    • Differential amplifiers are used to reject common-mode noise and amplify the difference between two input signals
    • Typical ECG amplifier gain is 1000-5000, with a bandwidth of 0.05-100 Hz
  • Filters remove unwanted frequency components from the ECG signal
    • High-pass filters remove low-frequency baseline wander (0.5-1 Hz cutoff)
    • Low-pass filters remove high-frequency noise and artifacts (50-100 Hz cutoff)
    • Notch filters remove power line interference (50 or 60 Hz)
  • Analog-to-digital converters (ADCs) sample and digitize the analog ECG signal for digital storage, processing, and analysis
    • Sampling rates of 250-1000 Hz are commonly used in ECG acquisition
    • Resolution of 12-16 bits is typical for ECG ADCs

Signal Processing in ECG

  • Signal processing techniques are used to improve the quality and interpretability of ECG signals
  • Filtering is used to remove noise and artifacts from the ECG signal
    • Digital filters (FIR, IIR) are implemented in software for more flexibility and control
    • Adaptive filters can adjust their parameters based on signal characteristics
  • Baseline wander correction removes low-frequency drift in the ECG signal
    • High-pass filtering, polynomial fitting, or cubic spline interpolation can be used
  • Powerline interference removal eliminates 50 or 60 Hz noise from the ECG signal
    • Notch filtering, adaptive filtering, or subtraction of a noise template can be used
  • QRS detection identifies the location and timing of QRS complexes in the ECG signal
    • Algorithms based on amplitude, slope, or wavelet transforms are commonly used
    • Enables the calculation of heart rate and the analysis of rhythm and morphology
  • Signal averaging improves the signal-to-noise ratio by combining multiple ECG beats
    • Reduces random noise and enhances small, consistent signal components (P wave, late potentials)
  • Time-frequency analysis (short-time Fourier transform, wavelet transform) provides information about the time-varying frequency content of the ECG signal
    • Useful for detecting and characterizing transient events and abnormalities

Clinical Applications of ECG

  • Diagnosis of cardiac arrhythmias
    • Identification of abnormal heart rhythms (bradycardia, tachycardia, atrial fibrillation, ventricular tachycardia)
    • Assessment of conduction disorders (heart block, bundle branch block)
  • Detection of myocardial ischemia and infarction
    • ST segment changes (elevation or depression) indicate reduced blood flow to the heart muscle
    • Q waves and T wave inversions can suggest previous myocardial infarction
  • Evaluation of electrolyte imbalances
    • Hyperkalemia can cause tall, peaked T waves and a prolonged PR interval
    • Hypokalemia can cause flattened or inverted T waves and a prolonged QT interval
  • Monitoring of drug effects on the heart
    • Antiarrhythmic drugs can prolong the QT interval and increase the risk of torsades de pointes
    • Digitalis toxicity can cause characteristic "scooped" ST segment depression
  • Risk stratification and prognosis
    • Left ventricular hypertrophy, as indicated by increased QRS voltage, is associated with increased cardiovascular risk
    • Prolonged QT interval is a risk factor for ventricular arrhythmias and sudden cardiac death
  • Guidance of therapy
    • Pacemaker and implantable cardioverter-defibrillator (ICD) placement and programming
    • Evaluation of the effectiveness of antiarrhythmic medications
    • Monitoring of cardiac resynchronization therapy (CRT) in heart failure patients

Challenges and Future Developments

  • Improving signal quality and reducing noise in ambulatory and long-term ECG monitoring
    • Development of more comfortable and less obtrusive wearable ECG devices
    • Use of active electrodes and shielding to minimize motion artifacts and interference
  • Enhancing the accuracy and reliability of automated ECG analysis algorithms
    • Machine learning and deep learning approaches for arrhythmia detection and classification
    • Incorporation of patient-specific information and context into analysis algorithms
  • Integrating ECG with other physiological signals and data sources
    • Multimodal monitoring that combines ECG with blood pressure, respiration, and activity data
    • Fusion of ECG with imaging data (echocardiography, cardiac CT, MRI) for comprehensive assessment
  • Developing personalized and predictive models for cardiovascular risk assessment
    • Using ECG features and other clinical data to stratify patients based on their risk profiles
    • Identifying early warning signs of impending cardiac events and guiding preventive interventions
  • Advancing telemedicine and remote ECG monitoring
    • Enabling real-time transmission and interpretation of ECG data from patients to healthcare providers
    • Facilitating timely diagnosis and management of cardiac conditions in underserved or remote areas
  • Exploring novel applications of ECG beyond the cardiovascular system
    • Using ECG-derived respiration (EDR) for respiratory monitoring
    • Analyzing heart rate variability (HRV) as a marker of autonomic function and stress
    • Investigating the potential of ECG in biometric identification and authentication


© 2024 Fiveable Inc. All rights reserved.
AP® and SAT® are trademarks registered by the College Board, which is not affiliated with, and does not endorse this website.

© 2024 Fiveable Inc. All rights reserved.
AP® and SAT® are trademarks registered by the College Board, which is not affiliated with, and does not endorse this website.